Changes in technique can reduce radiation dose in CT examination

Contact: Two minor changes in technique could make a major difference in the radiation dose used in survey CT scans, a new study shows. Survey scans are those that are performed before a regular CT scan, usually to plan for the examination.

The radiation dose in a survey scan can be as high as that of four chest X-rays, said Dianna D. Cody, PhD, associate professor in the department of imaging physics at The University of Texas, M.D. Anderson Cancer Center in Houston, TX. The radiation dose depends on X-ray tube kilovoltage (kVp), X-ray tube current and X-ray tube positioning, she said. Dr. Cody and her colleagues studied 21 CT scanners, representing three different vendors and 11 different models. They found that "the lowest radiation exposure was achieved using 80 kVp, minimum X-ray current and a 180-degree tube position. If these settings can be used for survey CT scans, the associated radiation exposure could be reduced to that of less than one chest X-ray," Dr. Cody said.

When the X-ray tube is put in the 180-degree position it is underneath the patient so the radiation beam strikes the table the patient is lying on first, "allowing the table to absorb the lowest-energy X-rays," she said. This reduces the exposure to the breast, in particular and does not have any effect on the quality of the CT survey image, Dr. Cody said.

On the other hand, "when the X-ray tube voltage or current is altered, the image quality will be affected," she said. "We encourage all radiologists to think about how the survey scan is used. If, for example, the survey CT scan is used only to plan for the examination, high-quality survey CT images may not be needed. In these cases we suggest that sites consider decreasing their X-ray tube voltage and current so that the relative radiation exposure from the survey scan is as low as possible, while maintaining a good enough image," she said.

"Default settings on CT scanners vary, and even thoug
'"/>


5-Aug-2005


Page: 1 2

Related biology news :

1. Changes in chromosomal constitution of preimplantation embryos suggest caution in genetic screening
2. Changes in west coast marine ecosystems significant
3. Changes in amino acids in the 1918 influenza virus cut transmission
4. Changes in dietary protein may override inherited skeletal abnormalities
5. Symposium Changes on Earth honors Yale Geologist Robert Berner
6. Changes to in utero environment may alter onset of cancer, U of T study
7. Changes in reef latitude
8. Changes to embryos can elicit change in adult fish
9. Changes in community size affect the outcome of competition
10. Innovative tagging technique may help researchers better protect fish stocks
11. Innovative research technique reveals another natural wonder in Yellowstone Park

Post Your Comments:
*Name:
*Comment:
*Email:


(Date:6/14/2017)... (NYSE: IBM ) is introducing several innovative partner startups ... collaboration between startups and global businesses, taking place in ... nine startups will showcase the solutions they have built with ... France is one of the ... percent increase in the number of startups created between 2012 ...
(Date:5/16/2017)... 16, 2017   Bridge Patient Portal , ... MD EMR Systems , an electronic medical ... GE, have established a partnership to build an ... the GE Centricity™ products, including Centricity Practice Solution ... These new integrations will allow healthcare delivery ...
(Date:4/19/2017)... New York , April 19, 2017 ... competitive, as its vendor landscape is marked by the ... the market is however held by five major players ... Safran. Together these companies accounted for nearly 61% of ... of the leading companies in the global military biometrics ...
Breaking Biology News(10 mins):
(Date:10/12/2017)... ... October 12, 2017 , ... ... platform specifically designed for life science researchers to analyze and interpret datasets, ... Franklin, who made a major contribution to the discovery of the double-helix ...
(Date:10/11/2017)... ... ... Personal eye wash is a basic first aid supply for any work environment, but most ... you rinse first if a dangerous substance enters both eyes? It’s one less decision, and ... unique dual eye piece. , “Whether its dirt and debris, or an acid or alkali, ...
(Date:10/11/2017)... , Oct. 11, 2017  VMS BioMarketing, a leading ... a nationwide oncology Clinical Nurse Educator (CNE) network, which will ... need for communication among health care professionals to enhance the ... nurses, office staff, and other health care professionals to help ... breast cancer. ...
(Date:10/11/2017)... ... October 11, 2017 , ... ... implantation and pregnancy rates in frozen and fresh in vitro fertilization (IVF) ... and maternal age to IVF success. , After comparing the results from the ...
Breaking Biology Technology:
Cached News: